What is the certified pH of RICCA RV150000-500A?
The nominal manufacturer specification
is pH 4.00 ± 0.01 at 25°C.
The exact assigned value depends
on the manufacturing lot.
Is this a Certified Reference Material?
Yes. RICCA identifies VeriSpec®
RV150000 as a CRM manufactured and
tested under ISO/IEC 17025
and ISO 17034 accreditation.
Is the buffer NIST-traceable?
RICCA's certification identifies
traceability to NIST SRM 185i,
186-I-g, and 186-II-g.
What is the actual pH for example lot 16042088?
The manufacturer reported pH 3.999
at 25°C. That result belongs to
lot 16042088 and must not
be assigned to another lot.
What measurement uncertainty applies?
The example certificate reports
an expanded uncertainty of 0.025 pH unit
with coverage factor k=2.
Review the certificate for the
actual received lot.
Why is the expanded uncertainty different from the ±0.01 pH specification?
The product specification and expanded
uncertainty describe different metrological
quantities. They must be considered
separately when evaluating the
reference material's suitability.
Is RV150000 color-coded red?
No. RV150000 is documented as
a colorless liquid. RICCA also offers
separately identified red
color-coded pH 4.00 buffers.
What chemical provides the buffer system?
Potassium hydrogen phthalate,
CAS 877-24-7, is the principal
buffering component.
Is a preservative present?
The RICCA SDS identifies a proprietary
preservative at less than 0.1% by weight.
Its exact chemical identity is not
provided in the reviewed SDS.
At what temperature is the pH certified?
RICCA certifies the assigned pH at 25°C.
Values published for other temperatures
are calculated rather than individually
measured and certified for each lot.
Should the bottle be mixed before use?
Yes. RICCA instructs users to mix
thoroughly by inversion before each use.
Can the CRM be diluted?
No. The manufacturer instructs
users not to dilute the reference buffer.
Can dispensed buffer be returned to the bottle?
No. Returning used buffer may contaminate
the stock reference material and compromise
its suitability for calibration.
Does opening the bottle affect the stated uncertainty?
RICCA's example certificate states
that the certified-value uncertainty
increases by 0.0021 pH unit for
each opening. Follow the applicable
manufacturer certificate and laboratory
uncertainty-control procedure.
What is the unopened shelf life?
RICCA specifies 24 months.
The actual expiration date must be
verified for the specific received lot.
How should the buffer be stored?
Recommended storage is 15-30°C.
Keep tightly closed, protect from
freezing and physical damage,
and avoid acid or base fumes.
Is the solution hazardous for transportation?
The reviewed SDS identifies the material
as not regulated under DOT, IATA,
and Canadian TDG dangerous-goods
transportation requirements.
Does the publicly available CoA certify every bottle sold?
No. The publicly available certificate
documents a specific lot. A laboratory
must verify the certificate corresponding
to the material actually received.
Is the product sterile or ISO cleanroom-classified?
No product-specific sterility or
ISO cleanroom compatibility certification
is established by the reviewed
manufacturer documents. Facility qualification
is necessary where controlled-environment
requirements apply.